Normal Parameters of Ear Acoustic Emissions in Neonates
Hernández HRJ, Hernández ALM, Castillo MN, De la Rosa MN, Martínez EJ, Flores SR, Torcida GME, Hernández NR
Language: Spanish
References: 15
Page: 63-68
PDF size: 116.85 Kb.
ABSTRACT
Introduction: the ear acoustic emissions (EAE) are produced from pure low intensity tones coming from the cochlea in response to a stimulus. Hypocusia in the neonate may be detected by universal hearing screening programs using EAE that have a sensitivity of 91 % and specificity of 85 %.
Objective: to define the normal limits we measured the EAE in a group of healthy newborns and calculated the media and standard deviation (SD). Patients and methods: We enrolled 280 healthy neonates, previous external ear evaluation for permeability and we measured the EAE at 7 to 30 days of life at 2000, 2500, 3187, 4000, 5062 y 6375 Hz, using a 55-65 dB stimulus for each frequency band.
Results: in each frequency band evaluated we calculated the media: 10.76, 9.16, 9.46, 7.89, 4.83, 2 dB SPL, and minus two SD: 0.54, –0.33, 0.06, –2.04, –4.37, –7.43 dB SPL, that we considered the lower limit of normality for both ears in each frequency band tested. There were no significant differences between measure-ments in both ears (p › 0.05).
Conclusions: the EAE measurement is an easy, and low cost test that can be used to make the universal hearing screening accessible. These results support the normal levels of EAE and the Corti organ function in healthy newborns.
REFERENCES
1. Kennedy CR. Neonatal screening for hearing impairment. Arch Dis Child 2000;83:377-382.
2. Bustos L. Evaluación auditiva. Apuntes de otorri-nolaringología. Disponible en http://escuela.med. puc.cl/paginas/publicaciones/Otorrino/Otorrino_ A011.html
3. Kennedy C, McCann D. Universal neonatal hearing screening moving from evidence to practice. Arch Dis Child 2004;89:F378-F383.
4. Johnson JL, White KR, Widen JE, Gravel JS, James M, Kennalley T, et al. A multicenter evaluation of how many infants with permanent hearing loss pass a two-stage otoacoustic emissions/automated auditory brainstem response newborn hearing screening protocol. Pediatrics 2005;116: 663-672.
5. Saurini P, Nola G, Lendvai D. Otoacoustic emissions: a new method for newborn hearing screening. Eur Rev Med Pharmacol Sci 2004;8: 129-133.
6. Silva A, Ysunza A, Berruecos P. Efecto del enmas-caramiento contralateral sobre latencia de las emisiones otoacústicas por productos de distorsión. An ORL Mex 1997;42:126-130.
7. Norton SJ, Gorga MP, Widen JE, Vohr BR, Folsom RC, Sininger YS, et al. Identification of neonatal hearing impairment: transient evoked otoacoustic emissions during the perinatal period. Ear Hear 2000;21:425-442.
8. DeCristofaro J, Bradley M, Claus-Parodi ST. Universal newborn hearing screening for congenital hearing loss: how much does it cost? Pediatr Res 1997;41:194.
9. Janssen T, Klein A, Gehr DD. Automated hearing threshold estimation in newborns using extrapolated DPOAE input/output functions. HNO 2003;51: 971-980.
Janssen T, Gehr DD, Klein A, Muller J. Distortion product otoacoustic emission for hearing threshold estimation and differentiation between middle-ear and cochlear disorders in neonates. J Acoust Soc Am 2005;117:2969-2979.
Hatzopoulus S, Pelosi G, Petruccelli J, Rossi M, Vigi V, Chierici R, et al. Efficient otoacoustic emission protocols employed in a hospital-based neonatal screening program. Acta Otolaryngol 2001;121:269-273.
Kezirian EJ, White KR, Yueh B, Sullivan SD. Cost and cost-effectiveness of universal screening for hearing loss in newborns. Otolaryngol Head Neck Surg 2001;124:359-367.
Matsumura M, Chida E, Suto S, Fukuda S, Kashiwamura M, Kuroda T, et al. Utility of OAE screener (GSI 70) for the evaluation of distortion product otoacoustic emissions. Nippon Jibiinkoka Gakkai Kaiho 2001;104:721-727.
Bitner GM. Hereditary deafness and phenotyping in humans. Br Med Bull 2002;63:73-94.
Steel K. New interventions in hearing impairment. BMJ 2000;320:622-625.