2012, Number 2
<< Back Next >>
Revista Cubana de Ortopedia y Traumatología 2012; 26 (2)
Spongy bone tissue sterilized with ethylene oxide gas
Jacas TMF, Sánchez NEO, García MNR, Piña AD
Language: Spanish
References: 34
Page: 143-155
PDF size: 75.50 Kb.
ABSTRACT
Objective: verify the sterilizing properties of ethylene oxide gas over the spongy bone used for implants in medical practice, mainly its efficacy against certain microorganisms and the time required for sterilization in industrial autoclaves.
Methods: 54 cube-shaped spongy bone tissue samples were prepared and distributed into 3 groups, each with 18 samples. The first group was used to evaluate a bacteriological contamination method with strains of known microorganisms. Upon contamination with the known microorganisms, the second and third groups were subjected to the effect of ethylene oxide gas in subgroups of 6 samples during 60, 90 and 120 minutes.
Results: in the first group 100 % of the samples contaminated were positive. In the second and third groups, the effectiveness of ethylene oxide gas was 100 % at 120 minutes.
Conclusion: this is a safe method appropriate for use in medical practice.
REFERENCES
Woodward LT. Les Grandes Etapes de la Chirurgie. París: Ed. Denoél; 1964.
Bado JL. Historia de la Ortopedia. Uruguay [citado 19 Sep 2006]. Disponible: http://www.smu.uy/historia/ortopedia/histOrtopedia.bado 1965.pdf
3.Álvarez Cambras R. Historia de la Cirugía Ortopédica y Traumatología Tratado de Ortopedia y Traumatología. T. 1 La Habana: Ed. Pueblo y Educación; 1985
Inclán A. The use of preserved bone graft in orthopaedic Surgery. J.Bone Joint Surg. 1942;24A:81.
Campbell W. Cirugía Ortopédica. Obras Completas. España: Ed. Harcourt; 2001.
RregulacionesCecmed007. Esterilización con óxido de etileno. Directrices sobre Buenas Prácticas. Regulación No. 16. La Habana:CECMED; 2000. p. 15.
Conclusiones II Congreso Panamericano de Esterilización. Perú: ANEEE; 2002.
Hernández Sampier R. Metodología de la Investigación. t 2. Cap. 10 . La Habana: Ed. Félix Varela; 2004. p. 343.
Esterilización con óxido de etileno. Circular Farmacéutica. 1980;38(julio-agosto-septiembre):268.
Comité Estatal de Normalización. Buenas Prácticas de Laboratorio, NC 26-212. La Habana: MINSAP; marzo 1992.
Ihoest W. L´oxide d´ethilene agend moderne de esterilization. J Pharm Belg. 1977;32(5):480-94.
Rendell B, Roberts RB. Safe use of ethylene oxide sterilization in hospitals. Anesth Analg Clev. 1970;49:919-21.
Guía para la salud y seguridad. No. 16 Óxido de Etileno. La Habana: OMS; 1993.
Jacas Torné M, Álvarez Cambras R, Aria Casson R, Sánchez Noda E. Freeze-dried and irradiated tricortical cancellous bovine grafts. A new product for the fussion to inter somatic cervical. Kuala Lumpur, Malaysia: 5th World Congress on Tissue Banking; 2008.
Jacas Torné M, Sánchez Noda E. Gas óxido de etileno. Una alternativa para la esterilización de tejidos implantables. Cienfuegos, Cuba: Congreso Ortopedia; 2008.
DL Zaragoza. Efecto de la radiación gamma sobre algunos injertos [citado 19 Nov 2011]. Disponible en: http://www.quimicanuclear.org
CCEEM. Boletín Informativo No. 2. Evaluación Preclínica Biológica de equipos médicos. No. 2 julio. La Habana: Centro de Control Estatal de Equipos Médicos; 1997.
Compuestos Heterocíclicos. La Habana: Memoria VII Congreso Sociedad Cubana de Bioingeniería; 2007.
Ibarra Fernández de la Vega CJ. Implantación de límites de la exposición ocupacional a sustancias nocivas en Cuba. Situación actual y perspectivas. Rev Cubana Hig Epidemiol. 1996;34(2):38-43.
OPS. Contaminación Ambiental en Centros Sanitarios. Rev Centro Salud. 1996;4(7):465-72.
Aragón Peña A, Gonzáles García MI. Protocolo de vigilancia sanitaria específica óxido de etileno. Galicia, España: Ministerio de Sanidad y Consumo; 2003.
Rutala WA, Wether DJ. New disinfection and sterilization methods. Energin Inf Dis. 2001;7(2):348-53.
Tarragano R, Cerdá N, López O, Valdés I. Esterilización con óxido de etileno. Bol AAM. 2004;163.
Rutala WA, Gergen MF, Wether DJ Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and paracetic acid. Am J Infect Control. 1998;26:393-8.
Philips CR, Kaye S. The sterilizing action of gaseous ethylene oxide. Am J Hygiene. 1949;50:270.
Hidalgo Rodríguez R, Castellanos Fernández V, Chiroles Despaigne S, Villavicencio Betancourt O. Dispositivos médicos de uso único reprocesados por esterilización química mediante óxido de etileno. Rev Cubana Hig Epidemiol. 2002;40(2):89-94.
Rhodes BS, Matthew W. Perspectives on reprocessing of single-use devices. J Clin Engineering. 2008;33(4):197-9.
O´Leary K, Watkins ED, Guess WL. Toxicological studies on certain medical grade plastics sterilized by ethylene oxide J Ph Sci. 1968;57:12.
WHO International Agency for Research on Cancer (IARC). Monographs on the evaluative of carcinogenic risks chemicals to men. Vol II. Geneva; Word Health Organization; 1976.
Rosell Farás G, Arias Castaño P. Óxido de etileno: prevención de la exposición en hospitales. España: Hospital Clínica I Provincia de Barcelona. Instituto Nacional de Seguridad e Higiene en el Trabajo; 2010.
Agency for toxic substances and disease registery (ATSDR)[citado 12 Sep 2008]. Avilable at: http://www.atsdr.cdc.gov/es/
Hospital Pediátrico William Soler. Actualizar las Políticas de Desinfección. Esterilización y antisepsia. Disponible en: http://www.sld.cu/sitios/william soler
Castellanos Fernández V, Rios Hernández M, Chiroles DespaigneS, Villavicencio Betancourt O, Hidalgo Rodríguez R. Residuales de la esterilización con gas óxido de etileno en equipos médicos. La Habana: Memorias V Congreso de la Sociedad Cubana de Bioingeniería; 2003.
Hidalgo Rodrigues R, Chiroles Despaigne S, Villavicencio Betancourt O. Evaluación cuantitativa de la eficacia de un esterilizador químico que emplea formaldehído 2 % en fase de vapor a bajas temperaturas. Rev Cubana Invest Bioméd [revista en la Internet]. 2006 Mar [citado 2012 Nov 23];25(1). Disponible en: http://scielo.sld.cu/scielo.php?script=sci_arttext&pid=S0864-03002006000100009&lng=e