Rev Mex Oftalmol 2009; 83 (4)
Profundidad de la cámara anterior mediante Ultrasonido A, Orbscan II, OCT visante y UBM. Estudio comparativo
Villanueva-Pérez G, Perdiz-Calvo L, Juárez-Flores P, Castañeda-García F
Language: Spanish
References: 12
Page: 221-225
PDF size: 184.12 Kb.
ABSTRACT
Purpose: To evaluate and compare the depth of the anterior chamber by means of ultrasound A (USG A), Corneal Topography (Orbscan II), Optical coherence Tomography (Visante OCT) and Ultrasonic biomicroscopy (UBM).
Material and methods: Prospective, comparative, longitudinal and observational study. Patients with cataract diagnosis without retinal vitreous pathology were included. In each patient an average of 5 measurements which each system were realized by the same person.
Results: Fifty eyes of 26 patients were studied, 50% of the female, with an average age of 69 years. A variance analysis was realized (ANOVA) with a value of p‹0.05 and a coefficient of correlation of 0, 97 between OCT versus UBM in comparison of 0, 78 between USG A Versus Orbscan.
Conclusions: The difference between systems is statistically significant. There is a greater correlation between the OCT and the UBM. The data collected by Orbscan were always smaller unlike the other systems. It is always necessary to evaluate the patient in an integral way and choose the best system available.
REFERENCES
Buehl W, Danijela S. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol 2006; 141:7-12.
Yun L, Jun-Heon K. Comparison of anterior chamber depth measurement between orbscan II and ultrasound biomicroscopy. J Refrac Surg 2007; (23).
Konstantopoulos A, Hossain P. Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis? Br J Ophthalmol 2007; 91:551-557.
Dada T, Sihota R. Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment. J Cataract Refract Surg 2007; 33:837-840.
Nemeth G, Vajas A. Assessment and reproducibility of anterior chamber depth measurement with anterior segment optical coherence tomography compared with immersion ultrasonography. J Cataract Refract Surg 2007; 33:443-447.
Nemeth J, Fekete O. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg 2003; 29:85-88.
Reddy A, Pande M. Comparative estimation of anterior chamber depth by ultrasonography, Orbscan II, and iol master. J Cataract Refract Surg 2004; 30:1268-1271.
Olsen T. Source of error in intraocular lens power calculation. J Cataract Refract Surg 1992; 18:125-129.
Solomon OD. Corneal indentation during ultrasonic pachometry. Cornea 1999; 18:214-215.
Vertugno M, Cardascia M. Anterior chamber depth measured by two methods in myopic and hyperopic phakic IOL implant. Br J Ophthalmol 2000; 84:1113-1116.
Auffarth GU, Tetz MR. Measuring anterior chamber depth with Orbscan topography system. J Cataract Refract Surg 1997; 23:1351-1355.
Pavlin CJ, Sherar MD. Ultrasound biomicroscopy of anterior segment structures in normal glaucomatous eyes. Am J Ophthalmol. 1992; 113:381-389.