Precision and accuracy of “a pocket” pulse oximeter in Mexico City
Torre-Bouscoulet L, Chávez-Plascencia E, Vázquez-García JC, Pérez-Padilla R
Language: Spanish
References: 17
Page: 28-33
PDF size: 90.39 Kb.
ABSTRACT
Background. Pulse oximeters are frequently used in the clinical practice and we must known their precision and accuracy. The objective was to evaluate the precision and accuracy of a “pocket” pulse oximeter at an altitude of 2,240 m above sea level.
Methods. We tested miniature pulse oximeters (Onyx 9,500, Nonin Finger Pulse Oximeter) in 96 patients sent to the pulmonary laboratory for an arterial blood sample. Patients were tested with 5 pulse oximeters placed in each of the fingers of the hand oposite to that used for the arterial puncture. The gold standard was the oxygen saturation of the arterial blood sample.
Results. Blood samples had SaO
2 of 87.2 ± 11.0 (between 42.2 and 97.9%). Pulse oximeters had a mean error of 0.28 ± 3.1%. SaO
2 = (1.204 x SpO
2) – 17.45966 (r = 0.92, p ‹ 0.0001). Intraclass correlation coefficient between each of five pulse oximeters against the arterial blood standard ranged between 0.87 and 0.99. HbCO (2.4 ± 0.6) did not affect the accuracy.
Conclusions. The miniature oximeter Nonin is precise and accurate at 2,240 m of altitude. The observed levels of HbCO did not affect the performance of the equipment. The oximeter good performance, small size and low cost enhances its clinical usefulness.
REFERENCES
Davidson I, Henry JB (Editores). Todd-Sanford, Clinical Diagnosis by laboratory methods. WB Saunders 1974.
American Association for Respiratory Care. AARC Clinical Practice Guideline: Pulse oximetry. Respir Care 1991; 36: 1406-9.
Pérez-Padilla R, Bracamonte-Peraza R, Manrique G, Ruíz-Primo ME. Exactitud del oxímetro de oído (Biox-III) en la Ciudad de México y su sensibilidad a la carboxihemoglobina. Arch Inst Cardiol Méx 1986; 56: 303-7.
Ries AL, Farrow JT, Clausen JL. Accuracy of two ear oximeters at rest and during exercise in pulmonary patients. Am Rev Resp Dis 1985; 132: 685-9.
Bickler PE, Feiner JR, Severinghaus JW. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology 2005; 102(4): 715-9.
Ralston AC, Webb RK, Runciman WB. Potential errors in pulse oximetry. III: Effects of interferences, dyes, dyshaemoglobins and other pigments. Anaesthesia 1991; 46(4): 291-5.
Douglas NJ, Brash JM, Wraight PK. Accuracy, sensitivity to carboxyhemoglobin and speed of response of the new Hewlett-Packard 47201A ear oximeter. Am Rev Resp Dis 1979; 119: 311-13.
Tashiro C, Koo YH, Fukumitsu K, Tomi K, Mashimo T, Yoshiya I. Effects of carboxyhemoglobin on pulse oximetry in humans. J Anesth 1988; 2: 36-40.
Trivedi NS, Ghouri AF, Lai E, Shah NK, Barker SJ. Pulse oximeter performance during desaturation and resaturation: a comparison of seven models. J Clin Anesth 1997; 9(3): 184-8.
Pérez-Padilla R, Torre-Bouscoulet L, Chávez E, Vázquez JC, Corcho A, Franco F, Muino A, Menezes AM, Márquez M, López MV. Prevalence of oxygen desaturation in adults at sea level and at an altitude of 2,240 m. Proc Am Thorac Soc 2005; 2: A778.
Perez-Padilla R. Population distribution residing at different altitudes: implications for hypoxemia. Arch Med Res 2002; 33: 162-6.
Beall CM, Almasy LA, Blangero J, Williams-Blangero S, Brittenham GM, Strohl KP, Decker MJ, Vargas E, Villena M, Soria R, Alarcón AM, Gonzales C. Percent of oxygen saturation of arterial hemoglobin among Bolivian Aymara at 3,900-4,000 m. Am J Phys Anthropol 1999; 108: 41-51.
Wimberley PD, Siggaard-Andersen O, Fogh-Andersen N, Zijlstra WG, Severinghaus JW. Haemoglobin oxygen saturation and related quantities: definitions, symbols and clinical use. Scand J Clin Lab Invest 1990; 50(4): 455-9.
Kramer MS, Feinstein AR. Clinical biostatistics. LIV. The biostatistics of concordance. Clin Pharmacol Ther 1981; 29(1): 111-23.
STATATM version 8.0, Stata Co, Texas, USA.
Medical Research Council Working Party. Long-term domiciliary oxygen therapy in chronic hypoxic cor pulmonale complicating chronic bronchitis and emphysema. Lancet 1981; 1: 681-6.
Nocturnal Oxygen Therapy Trial Group. Continuous or nocturnal oxygen therapy in hypoxemic chronic lung disease: a clinical trial. Ann Intern Med 1980; 93: 391-8.