2019, Number 1
<< Back
Biotecnol Apl 2019; 36 (1)
Strong impact of the side scatter channel performance in the volumetric flow cytometer CD4+ T cell counts using the CD4% easy count kit
Iglesias E, Hernández-Requejo D, Godínez-López MC, Vazquez-Arteaga A
Language: English
References: 24
Page: 1301-1306
PDF size: 827.53 Kb.
ABSTRACT
The absolute number and percentage of CD4+ T cells are important parameters in the follow up of HIV-1-infected patients. However, discordant results were found between the CD4+ cell counts and percentage of cells, determined by flow cytometry using the CD4% easy count kit from Sysmex-Partec®. This is a validated kit based on a single-platform methodology for volumetric cytometry counting, and deviations were present despite quality controls were in the acceptable range. Therefore, the photomultiplier performance was analyzed as the possible source of deviation, using pre-counted beads, for the individual channels: side scatter (SSC), fluorescence 1 (FL-1) and FL-2 used in the CD4% easy count kit. Additionally, forward scatter (FSC) and FL-3 were selected for comparison. Then, the coefficients of variation (CV) and accuracy were calculated and beads concentration (count/mL) obtained within the peak of histograms in every optical channel and compared by linear regression analysis. CV, accuracy and linear regression analysis showed fine goodness-of-fit for all channels, but the slope of the SSC was shown statistically different (p < 0.01), suggesting bad performance on that channel. Other minor deviations were recorded comparing the intercept of the FSC to FL-1 (p = 0.03) and FL-3
(p = 0.02). Our results explained previous discordances using the CD4% easy count kit and revealed a previously unappreciated limitation of routine quality controls. They provide new routine quality controls which can be incorporated to detect photomultiplier’s error in any flow cytometer even without significant deterioration of performance.
REFERENCES
WHO, editor. Consolidated guidelines on the use of antiretroviral drugs for treating and preventing hiv infection. Recommendations for a public health approach. 2nd ed. Geneva: WHO Library Cataloguing-in-Publication Data; 2016.
Panel on Opportunistic Infections in HIVInfected Adults and Adolescents. Guidelines for the prevention and treatment of opportunistic infections in HIV-infected adults and adolescents: recommendations from the Centers for Disease Control and Prevention, the National Institutes of Health, and the HIV Medicine Association of the Infectious Diseases Society of America. 2017 [cited: 2019 Jan 17]. Available from: http://aidsinfo.nih.gov/contentfiles/ lvguidelines/adult_oi.pdf
Brannstrom J, Svedhem Johansson V, Marrone G, Wendahl S, Yilmaz A, Blaxhult A, et al. Deficiencies in the health care system contribute to a high rate of late HIV diagnosis in Sweden. HIV Med. 2016;17(6):425-35.
Drain PK, Losina E, Parker G, Giddy J, Ross D, Katz JN, et al. Risk factors for latestage HIV disease presentation at initial HIV diagnosis in Durban, South Africa. PLoS One. 2013;8(1):e55305.
Glencross D, Scott LE, Jani IV, Barnett D, Janossy G. CD45-assisted PanLeucogating for accurate, cost-effective dual-platform CD4+ T-cell enumeration. Cytometry. 2002;50:69-77.
Dieye TN, Vereecken C, Diallo AA, Ondoa P, Diaw PA, Camara M, et al. Absolute CD4 T-cell counting in resource-poor settings: direct volumetric measurements versus bead-based clinical flow cytometry instruments. J Acquir Immune Defic Syndr. 2005;39(1):32-7.
Koyalta D, Jenabian MA, Djouater B, Djemadji-Oudjeil N, Mbopi-Keou FX, Ndjoyi-Mbiguino A, et al. Field evaluation in Chad of community usage of CD4 T lymphocyte counting by alternative singleplatform flow cytometry. BMC Health Serv Res. 2013;13:373.
Pattanapanyasat K, Lerdwana S, Noulsri E, Chaowanachan T, Wasinrapee P, Sakulploy N, et al. Evaluation of a new single-parameter volumetric flow cytometer (CyFlow(green)) for enumeration of absolute CD4+ T lymphocytes in human immunodeficiency virus type 1-infected Thai patients. Clin Diagn Lab Immunol. 2005;12(12):1416-24.
Pattanapanyasat K, Phuang-Ngern Y, Sukapirom K, Lerdwana S, Thepthai C, Tassaneetrithep B. Comparison of 5 flow cytometric immunophenotyping systems for absolute CD4+ T-lymphocyte counts in HIV-1-infected patients living in resourcelimited settings. J Acquir Immune Defic Syndr. 2008;49:339-47.
Thairu L, Katzenstein D, Israelski D. Operational challenges in delivering CD4 diagnostics in sub-Saharan Africa. AIDS Care. 2011;23:814-21.
Glencross DK, Aggett HM, Stevens WS, Mandy F. African regional external quality assessment for CD4 T-cell enumeration: development, outcomes, and performance of laboratories. Cytometry B Clin Cytom. 2008;74 Suppl 1:S69-79.
Koyalta D, Jenabian MA, Nadjiouroum N, Djouater B, Djemadji-Oudjeil N, Ndjoyi- Mbiguino A, et al. Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resourceconstrained settings. BMC Res Notes. 2013;6:169.
WHO. Prequalification of in vitro diagnostics public report. Product: CyFlow® Counter System with CD4 easy count kit and CD4% easy count kit. WHO reference number: PQDx 0350-081-00; Geneva: WHO; 2018.
Hulgan T, Raffanti S, Kheshti A, Blackwell RB, Rebeiro PF, Barkanic G, et al. CD4 lymphocyte percentage predicts disease progression in HIV-infected patients initiating highly active antiretroviral therapy with CD4 lymphocyte counts >350 lymphocytes/mm3. J Infect Dis. 2005;192(6):950-7.
Hulgan T, Shepherd BE, Raffanti SP, Fusco JS, Beckerman R, Barkanic G, et al. Absolute count and percentage of CD4+ lymphocytes are independent predictors of disease progression in HIV-infected persons initiating highly active antiretroviral therapy. J Infect Dis. 2007;195(3):425-31.
Pol S, Artru P, Thepot V, Berthelot P, Nalpas B. Improvement of the CD4 cell count after alcohol withdrawal in HIV-positive alcoholic patients. AIDS. 1996;10:1293-4.
Bloemena E, Weinreich S, Schellekens PT. The influence of prednisolone on the recirculation of peripheral blood lymphocytes in vivo. Clin Exp Immunol. 1990;80:460-6.
Mili F, Flanders WD, Boring JR, Annest JL, Destefano F. The associations of race, cigarette smoking, and smoking cessation to measures of the immune system in middle-aged men. Clin Immunol Immunopathol. 1991;59:187-200.
Hull MW, Rollet K, Odueyungbo A, Saeed S, Potter M, Cox J, et al. Factors associated with discordance between absolute CD4 cell count and CD4 cell percentage in patients coinfected with HIV and hepatitis C virus. Clin Infect Dis. 2012;54(12):1798-805.
Laurence J. T-cell subsets in health, infectious disease, and idiopathic CD4+ T lymphocytopenia. Ann Intern Med. 1993;119:55-62.
Whitby L, Whitby A, Fletcher M, Barnett D. Current laboratory practices in flow cytometry for the enumeration of CD 4(+) T-lymphocyte subsets. Cytometry B Clin Cytom. 2015;88:305-11.
Rigolin GM, Lanza F, Castoldi G. Photomultiplier voltage setting: possible important source of variability in molecular equivalents of soluble fluorochrome (MESF) calculation? Cytometry. 1995;20:362-8.
Peeling RW, Sollis KA, Glover S, Crowe SM, Landay AL, Cheng B, et al. CD4 enumeration technologies: a systematic review of test performance for determining eligibility for antiretroviral therapy. PLoS One. 2015;10(3):e0115019.
Lawrie D, Coetzee LM, Glencross DK. A model for continuous quality control incorporating sample-to-sample assessment of optical alignment, fluorescence sensitivity, and volumetric operation of flow cytometers. Cytometry B Clin Cytom. 2010;78:201-10.