Acute dislocation of a Perceval valve treated with TAVI valve-in-valve: a literature review and case report
Morales-Portano JD, Zaldívar-Fujigaki JL, García-García JF, Madrid-Dour EA, Muratalla-González R, Merino-Rajme JA
Language: English
References: 16
Page: 56-60
PDF size: 226.46 Kb.
ABSTRACT
Dislocation of a Perceval valve is an extremely rare situation, with only a handful of proven cases entailing migration or dysfunction of the valve. A 76-year-old male with severe dyspnea with a severe aortic stenosis revealed by transthoracic echocardiography (TTE); surgical procedure with a Perceval valve was made without any complications. Within the first 24-hours in the Intensive Care Unit, the patient presented acute heart failure. The transesophageal echocardiography (TEE) showed severe paravalvular leak, then a fluoroscopy confirmed the dislocated and dysfunctional Perceval valve. An emergency transcatheter aortic valve-in-valve procedure with a CoreValve 29 mm was performed to correct the dislocated valve. To the best of our knowledge, this is the first documented case in the world of an acute dislocated Perceval valve treated with a TAVI valve-in-valve.
REFERENCES
Cowell S, Newby D, Boon N, et al. Calcific aortic stenosis: same old story? Age and Ageing. 2004; 33: 538-544.
Iung B, Baron G, Butchart EG et al. A prospective survey of patients with valvular heart disease in Europe: The Euro Heart Survey on Valvular Heart Disease. Eur Heart J. 2003; 24: 1231-1243.
Shrestha M, Folliguet T, Meuris B, Dibie A et al. Sutureless Perceval S aortic valve replacement: a multicenter, prospective pilot trial. J Heart Valve Dis. 2009; 18: 698-702.
Zannis K, Folliguet T, Laborde F. New sutureless aortic valve prosthesis: another tool in less invasive aortic valve replacement. Curr Opin Cardiol. 2012; 27: 125-129.
Fischlein T, Meuris B, Hakim-Meibodi K et al. The sutureless aortic valve at 1 year: A large multicenter cohort study. J Thorac Cardiovasc Surg. 2016; 151: 1617-1626.
PARTNER 2 Investigators. Transcatheter or surgical aortic-valve replacement in intermediate-risk patients. N Engl J Med. 2016; 374: 1609-1620.
Loyalka P, Montgomery KB, Nguyen TC et al. Valve-in-valve transcatheter aortic valve implantation: a novel approach to treat paravalvular leak. Ann Thorac Surg. 2017; 104: e325-e327.
Dvir D, Webb JG, Bleiziffer S et al. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves. JAMA. 2014; 312: 162-170.
Shrestha M, Fischlein T, Meuris B et al. European multicenter experience with the sutureless Perceval valve: clinical and haemodynamic outcomes up to 5 years in over 700 patients. Eur J Cardiothoracic Surg. 2016; 49: 234-241.
Durand E, Tron C, Eltchaninoff H. Emergency transcatheter aortic valve implantation for acute and early failure of sutureless Perceval aortic valve. Can J Cardiol. 2015; 31: 1204.e13-5.
Mangner N, Holzhey D, Misfeld M et al. Treatment of a degenerated sutureless Sorin Perceval® valve using an Edwards SAPIEN 3. Interact Cardiovasc Thorac Surg. 2018; 26: 364-366.
Concistre G, Miceli A, Chiaramonti F et al. Delayed dislocation of a sutureless aortic bioprosthesis: the first case. Interact Cardiovasc Thorac Surg. 2012; 14: 892-893.
Amr G, Ghoneim A, Giraldeau G et al. First case of a sutureless Perceval valve delayed proximal migration. J Thorac Cardiovasc Surg. 2017; 153: e21-e23.
Cerillo AG, Amoretti F, Mariani M et al. Increased gradients after aortic valve replacement with the perceval valve: the role of oversizing. Ann Thorac Surg. 2018; 106: 121-128.
Baert J, Astarci P, Noirhomme P, de Kerchove L. The risk of oversizing with sutureless bioprosthesis in small aortic annulus. J Thorac Cardiovasc Surg. 2017; 153: 270-272.
Kalra A, Reyes M, Yang EY et al. Transcatheter aortic valve replacement for perceval sutureless aortic valve failure. J Invasive Cardiol. 2017; 29: E65-E66.