ABSTRACT
The pulmonary artery catheter (CAP) is a device used in intensive care units (ICUs) to measure pressures in the heart and pulmonary blood vessels as part of hemodynamic monitoring primarily in cardiac surgery patients. The USCOM device is a non-invasive technique that uses Doppler technology to obtain measurements of systolic volume and its derivatives.
The following CO measurement comparison is performed between these two devices in patients with septic shock.
An observational, prospective, longitudinal and comparative study was conducted in patients with septic shock aged between 18 and 60 years admitted to intensive care in the period May-June 2017.
In view of the decrease in the use of the pulmonary artery catheter due to the controversy of not improving the mortality in the patients of the Intensive Care Units, the placement of this catheter has fallen into disuse; however, cardiac output measured by the Swan Ganz catheter remains the «gold standard» for real-time measurement of cardiac output and systemic and pulmonary resistance.
The CO measurement by PAC versus USCOM correlates, in such a way, that USCOM measurement can be used in a patient with septic shock, who is not expected to invade to determine their hemodynamic conditions.
REFERENCES
Mateu CM, Ferrándiz SA, Gruartmoner VG, Mesquida FJ, Sabatier CC, Poveda HY y cols. Técnicas disponibles de monitorización hemodinámica. Ventajas y limitaciones. Med Intensiva. 2012;36(6):2010-2012.
García X, Mateu L, Maynar J, Mercadal J, Ochagavía A, Ferrandiz A. Estimación del gasto cardiaco. Utilidad en la práctica clínica. Monitorización disponible invasiva y no invasiva. Med Intensiva. 2011;35(9):552-561.
Wong LS, Yong BH, Young KK, Lau LS, Cheng KL, Man JS, et al. Comparison of the USCOM ultrasound cardiac output monitor with pulmonary artery catheter thermodilution in patients undergoing liver transplantation. Liver Transpl. 2008;14(7):1038-1043.
Tan HL, Pinder M, Parsons R, Roberts B, van Heerden PV. Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit. Br J Anaesth. 2005;94(3):287-291.
Carrillo LA, Fiol SM, Rodríguez SA. El papel del catéter de Swan-Ganz en la actualidad. Med Intensiva. 2010;34(3):203-214.
Ospina-Tascón GA, Cordioli RL, Vicent JL. What type of monitoring has been shown to improve outcomes in acutely ill patients? Intensive Care Med. 2008;34(5):800-820.
Hodgson LE, Forni LG, Venn R, Samuels TL, Wakeling GH. A comparison of the non-invasive ultrasonic cardiac output monitor (USCOM) with the oesophageal Doppler monitor during major abdominal surgery. J Intensive Care Soc. 2016;17(2):103-110.
Vincent JL, De Backer D. Circulatory shock. N Engl J Med. 2013;369(18):1726-1734.
García X, Mateu L, Maynar J, Mercadal J, Ochagavía A, Ferrandiz A. Estimating cardiac output. Utility in the clinical practice. Available invasive and non-invasive monitoring. Med Intensiva. 2011;35(9):552-561. doi: 10.1016/j.medin.2011.01.014.
Prys-Roberts C. The measurement of cardiac output. Br J Anaesth. 1969;41(9):751-760.
Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK. Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg. 2010;110(3):799-811.
Kadota LT. Theory and application of thermodilution cardiac output measurement: a review. Heart Lung. 1985;14(6):605-616.
Stewart GN. Researches on the Circulation Time and on the Influences which affect it. J Physiol. 1897;22(3):159-183.
Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D. Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med. 1970;283(9):447-451.
Nishikawa T, Dohi S. Errors in the measurement of cardiac output by thermodilution. Can J Anaesth. 1993;40(2):142-153.
Pinsky MR. Hemodynamic profile, interpretation. In: Tobin MJ, editors. Principles and practice of intensive care monitoring. New York: McGraw-Hill; 1998. p. 871-888.
Vincent JL, Bihari D, Suter PM, Bruining HA, White J, Nicolas-Chanoin MH, et al. The prevalence of nosocomial infection in intensive care units in Europe, Results of the European Prevalence of Infection in Intensive Care (EPIC) study. EPIC International Advisory Committee. JAMA. 1995;274(8):639-644.
Sakr Y, Vincent JL, Reinhart K, Payen D, Wiedermann CJ, Zandstra DF, et al. Use of the pulmonary artery catheter is not associated with worse outcome in the Intensive Care Unit. Chest. 2005;128(4):2722-2731.
Iberti TJ, Daily EK, Leibowitz AB, Panacek EA, Silverstein JH, Albertson TE. A multicenter study of physicians knowledge of the pulmonary artery catheter, Pulmonary Artery Catheter Study Group. JAMA. 1990;264(22):2928-2932.
Pölönen P, Ruokonen E, Hippeläinen M, Pöyhönen M, Takala J. A prospective, randomized study of goal oriented hemodynamic therapy in cardiac surgical patients. Anaesth Analg. 2000;90(5):1052-1059.
Tuchschmidt J, Fried J, Astiz M, Rackow E. Elevation of cardiac output and oxygen delivery improves outcome in septic shock. Chest. 1992;102(1):216-220.
Lobo SMA, Salgado PF, Castillo VG, Borim AA, Polachini CA, Palchetti JC, et al. Effects of maximizing oxygen delivery on morbidity and mortality in high risk surgical patients. Crit Care Med. 2000;28(10):3396-3404.
Kern JW, Shoemaker WC. Meta-analysis of hemodynamic optimization in high-risk patients. Crit Care Med. 2002;30:1686-1692.
Hadian M, Pinsky MR. Evidence-based review of the use of the pulmonary artery catheter: impact data and complications. Crit Care. 2006;10(Suppl 3):S8.
Boyle M, Lawrence J, Belessis A, Murgo M, Shehabi Y. Comparison of dynamic measurements of pulse contour with pulsed heat cpontinuous cardiac output in postoperative cardiac surgical patients. Aust Crit Care. 2007;20(1):27-32.
USCOM Ltd. Available in: [http://www.uscom.com.au/]. Australia: The Association; c2000-2009 [Updated 2009; cited 2016 Nov 24]. USCOM Ltd; [About 55 screens]. Available in: http://www.talentmed.com.tw/download/UscomBasics.pdf