2005, Number 4
<< Back Next >>
Bioquimia 2005; 30 (4)
Preparation and evaluation of a glucose determination kit (glucose oxidase/peroxidase)
Pérez-Espinoza M, Brambila E
Language: Spanish
References: 28
Page: 110-117
PDF size: 81.62 Kb.
ABSTRACT
In this work we describe the elaboration and evaluation of a glucose kit based in the Trinder method. Glucose kit evaluation included studies of linearity, within-run and between-day precision assays, recovery, and method comparison to determine total error of the kit. Results showed that the kit was linear up to 400 mg/dL, in addition it had an excellent precision with within-run and between-day coefficients of variation of 0.67% and 1.13%, respectively. Glucose kit was affected only by high haemoglobin and bilirubin concentrations, and high levels of ascorbic acid. Addition of several glucose amounts to the samples was 100% recovered. Method comparison studies using the hexocinase method as a reference method shown a highly correlation (r = 0.996), but, results were approximately 5 mg/dL lower in the glucose kit prepared in our laboratory. Total error in our method (5.8 mg/dL) was acceptable as compared with the maximum allowable error to glucose assays (10 mg/dL). Reference limits obtained from selected individuals were 67 to 111 mg/dL. Based in these results we state that the glucose kit prepared in our laboratory has the analytic quality to be used for plasma glucose determinations.
REFERENCES
Rodríguez-de-Romo AC. Una nueva forma de entender la enfermedad en el siglo XIX. Laborat-acta 2001; 13: 61-67.
Benedict SR. The detection and estimation of reducing sugars. J Biol Chem 1907; 3: 101-117.
Folin O, Wu H. A system of blood analysis. J Biol Chem 1919; 38: 147-159.
Nelson N. A photometric adaptation of the Somogyi method for the determination of glucose. J Biol Chem 1944; 153: 375-380.
Hoffman WS. A rapid photoelectric method for the determination of glucose in blood and urine. J Biol Chem 1937; 120: 51-55.
Skeggs LT Jr. Persistent and prayer: from the artificial kidney to the autoanalyzer. Clin Chem 2000; 46: 1425-1436.
Richterich R, Colombo JP. Hidratos de carbono. En: Química Clínica Teoría, práctica e interpretación. Barcelona: Salvat Editores: 1983. p. 304-321.
Hultman E. Rapid specific method for determination of aldosaccharides in body. Nature Lond 1959; 183: 108-109.
Abraham CV. A modified o-toluidine reagent for glucose analysis. Clin Chim Acta 1976; 70: 209-211.
Dietzler DN; Smith CH. Hidratos de carbono. En: Gradwohl. Métodos y diagnósticos del laboratorio clínico. 8ª ed. Buenos Aires: Editorial Médica Panamericana: 1983. p.190-225.
Dobrick LA. Screening method for glucose of blood serum utilizing glucose oxidase and an indophenol indicator. J Biol Chem 1958; 231: 403-409.
Passey RB, Gillum RL, Fuller JB, Urry FM, Giles ML. Evaluation and comparison of 10 glucose methods and the reference method recommended in the proposed product class standard 1974. Clin Chem 1977; 23: 131-139.
Lott JA, Turner K. Evaluation of Trinder’s glucose oxidase method for measuring glucose in serum and urine. Clin Chem 1975; 21: 1754-1760.
Westgard JO, Quam EF, Barry PL, Ehrmeyer SS. Method validation. 2nd. ed. Madison: Editorial Westgard Quality Corporation: 2003.
Westgard JO, Hunt MR. Use and interpretation of common statistical tests in method-comparison studies. Clin Chem 1973; 19: 49-57.
PetitClerk C, Wilding P. The theory of reference values. Part 2. Selection of individuals for the production of reference values. J Clin Chem Clin Biochem 1984; 22: 203-208.
Solberg HE. The theory of reference values. Part 5. Statistical treatment of collected reference values. Determination of reference limits. J Clin Chem Clin Biochem. 1983; 21: 749-760.
Sacks DB, Bruns DE, Goldstein DeMacLaren NK, McDonald JM, Parrot M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002; 48: 436-472.
Report of the expert committee on the diagnosis and classification of diabetes mellitus. Diabetes Care 2003; 26 (Suppl 1): S5-S20.
Sonowane M, Savory J, Cross RE, Heintges MG, Chester B. Kinetic measurement of glucose with a centrifugal analyzer; hexokinase and glucose oxidase procedures compared. Clin Chem 1976; 22: 1100-1101.
Gochman N, Schmitz JM. Application of a new peroxide indicator reaction to the specific, automated determination of glucose with glucose oxidase. Clin Chem 1972; 18: 943-950.
Witte DL, Brown LF, Feld RD. Effects of bilirubin on detection of hydrogen peroxide by use of peroxidase. Clin Chem 1978; 24: 1778-1782.
Maguire GA, Price CP. Evidence for interference by ascorbate in the measurement of cerebrospinal fluid glucose by a kinetic glucose oxidase/peroxidase procedure. Clin Chem 1983; 29: 1810-1812.
Lutz RA, Flückiger J. Kinetic determination of glucose with the GEMSAEC (ENI) centrifugal analyzer by the glucose dehydrogenase reaction, and comparison with two commonly used procedures. Clin Chem 1975; 21: 1372-1377.
Romano AT. Automated glucose methods: evaluation of a glucose oxidase-peroxidase procedure. Clin Chem 1973; 19: 1152-1157.
Westgard JO. Basic Planning for Quality. Appendix 1: CLIA 88 Analytical Quality Requirements. Madison: Westgard QC, Inc.; 2000: 245-248.
Walker RE, Bachorik PS, Kwiterovich PO. Evaluation of five enzymic kits for determination of triglyceride concentrations in plasma. Clin Chem 1982; 28: 2299-2304.
Solberg HE. The theory of reference values. Part 1. The concept of reference values. J Clin Chem Clin Biochem 1987; 25: 337-342.