2016, Number 2
Next >>
Rev Mex Ing Biomed 2016; 37 (2)
Respiratory Rate Detection by a Time-BasedMeasurement System
Sifuentes E; Cota-Ruiz J; González-Landaeta R
Language: English
References: 20
Page: 91-99
PDF size: 0. Kb.
ABSTRACT
This paper proposes a system that converts a time-modulated signal from a resistive sensor into a digital signalwith the goal to estimate the respiratory rate of a subject. To detect breathing, a known method based on a nasalthermistor, which detects temperature changes near the nostrils, is used. In this work, the thermistor mountedin a Wheatstone bridge, forms a
RC circuit which is connected directly to a microcontroller, without using anyanalog circuit or analog-digital converter. Thus, whenever the subject breathes, it causes a fractional change inresistance x(
ΔR/R0) on the thermistor, and this produces a time-modulated signal that is directly digitized withthe microcontroller. Measurements were made on 23 volunteers, obtaining changes ofx ›0.01. The temperatureresolution was 0.2
◦C, and the time response was 0.8 s, mainly limited by the thermistor properties; these featureswere enough to obtain a well-defined waveform of the breathing, from which was easy to estimate the respiratoryrate by a compact, low cost and low power consumption system. Unlike interface circuits based on voltage or currentamplitude, with this kind of interface, the self-heating of the sensor is avoided since the thermistor does not requireany voltage or bias current.
REFERENCES
Schäfer A., et al., “Estimation ofbreathing rate from respiratory sinusarrhythmia: comparison of variousmethods,”Ann. of Biomed. Eng., vol.36, no 3, pp. 476-485, Jan. 2008.
Karlen W., et al., “Multiparameterrespiratory rate estimation from thephotoplethysmogram,”IEEE Trans. onBiomed. Eng., vol. 60, no 7, pp. 1946-1953, Feb. 2013.
Chan A. M., et al., “Ambulatoryrespiratory rate detection using ECG and triaxial accelerometer,” in35thConf. of the IEEE Eng. in Med. andBiol. Soc., Osaka, 2013, pp. 4058-4061.
Carskadon M. A., et al., “Respirationduring sleep in children,”WesternJournal of Medicine, vol. 128, pp. 477-481, Jun. 1978.
American Thoracic Socienty “Standardsand indications for cardiopulmonarysleep studies in children,”Am. J Respir.Crit. Care Med., vol. 153, no. 2, pp. 866-878, Feb. 1996.
Dinç A. E., et al, “Reliability of SleepStrip as a screening test in obstructivesleep apnea patients,”Eur. Arch. ofOtorhinolaryngol., vol. 271, no. 10, pp.1-6, Oct. 2014.
Tan H. L., et al., “OvernightPolysomnography versus RespiratoryPolygraphy in the Diagnosis of PediatricObstructive Sleep Apnea,”Sleep, Vol.37, no. 2, pp. 255-260, Feb. 2014.
Pallàs-Areny R., Webster J. G.,“Sensors and Signal Conditioning 2ndEdition”, John Wiley & Sons, NewYork, 2001, pp. 94-109.
Cuadras A., Casas O., “Determinationof heart rate using a high-resolutiontemperature measurement,”IEEESensors J., Vol. 6, no. 3, pp. 836-843,Jun. 2006 .
Sifuentes E., et al., “Direct interfacecircuit to linearise resistive sensorbridges,”Sensors and Actuators A, vol.147, no. 1, pp. 210-215, Sep. 2008.
Jordana J., Pallàs-Areny R., “Asimple, efficient interface circuit forpiezoresistive pressure sensors,”Sensorsand Actuators A, vol 127, no. 1, pp. 69-73, Feb. 2006.
Sifuentes E., et al., “Wireless magneticsensor node for vehicle detection withoptical wake-up,”IEEE Sensors J., vol.11, no. 8, pp. 1669-1676, Jan. 2011.
Brown B. H., et al., “Medical Physicsand Biomedical Engineering,” Instituteof Physics, Madison, 2001, pp. 548-577.
Toba E., et al., “Non-InvasiveMeasurement System for HumanRespiratory Condition and BodyTemperature,” inIEEE Int. Conf. onMFI, Las Vegas NV, 1994, pp. 770-783.
Storck K., et al., “Heat TransferEvaluation of the Nasal ThermistorTechnique,”IEEE Trans. On Biomed.Eng., vol. 43, no. 12, pp. 1187-1191, Dec.1996.
Farre R., et al., “Accuracy ofthermistors and thermocouples asflow-measuring devices for detectinghypopnoeas,”Eur. Respir. J., vol. 11,no.1, pp. 179-182, Jan. 1998.
Sifuentes E., et al., “Improved directinterface circuit for resistive full-andhalf-bridge sensors,” inIEEE, 4th Int.Conf. on Electrical and ElectronicsEngineering, ICEEE 2007, Mexico,2007, pp. 197-200.
Reverter F., “The art of directlyinterfacing sensors to microcontrollers,”Low Power Electronics and ApplicationsJ., vol 2, no. 4, pp 265-281, 2012.
Reverter F., “Direct interface circuitsfor sensors”, in: S. Nihtianov and A.Luque, “Smart Sensors and MEMS:Intelligent Devices and Microsystemsfor Industrial Applications”, WoodheadPublishing, Chap 2, pp 27-62, 2014.
20.LDT1-028K Shielded Piezo Sensors,Measurement Specialties, Hampton,VA, 2009, pp. 1-3