The Six Sigma model in the Hematology internal quality control assessment
León PF
Language: Spanish
References: 15
Page: 5-8
PDF size: 322.83 Kb.
ABSTRACT
Six Sigma is a quality control model that was first introduced into the Industry in 1979 by Motorola Company. With this model possible causes for error can be detected so they're eliminated to improve the process. Six Sigma can be used for Internal Quality Control assessment in a regular basis for evaluation and joint with the External Quality Control in the integral analytic phase evaluation too. We made the six months sigma average calculation of Hematology Internal Quality Control, and used CLIA (Clinical Laboratory Improvement Amendments), SEQC (Sociedad Española de Química Clínica), Tonks, Aspen and Qualitat suggested specifications. The sigma average value of three quality control levels is 5.967 (CLIA) which is an optimal quality level.
REFERENCES
Ricós C, Perich C, Álvarez V, Biosca C. Aplicación del modelo Seis-Sigma a la mejora de la calidad analítica del laboratorio clínico. Rev Lab Clin 2009; 2(1): 2–7.
Terrés A. SIX SIGMA: determinación de metas analíticas. RevMex Patol Clin 2007; 54(1): 28–39.
Terrés A. Estimación de la incertidumbre y de la variabilidad total en el laboratorio clínico. Rev Mex Patol Clin 2006; 53(4): 185-196.
Sunil K., Lopamudra R. Quantitative application of Sigma metrics in medical biochemistry. J Clin Diagn Res 2013; 7(12): 2689-2691.
Terrés A. Mejorar la calidad al nivel Six Sigma integrando los resultados de la evaluación externa con los del programa interno aplicando el método QQCDC. Rev Mex Patol Clin 2010; 57(3): 110-121.
Westgard J, Westgard S. The quality laboratory testing today. Am J Clin Pathol 2006; 125: 343–354.
Norma ISO/IEC 15189. Laboratorios clínicos requisitos particulares para la calidad y la competencia. 2007.
Koch D, Jeffrey J, Quam E, Dowd D. Selection of medically useful quality-control procedures for individual test done in a multitest analytic system. Clin Chem 1990; 36(2): 230-233.
Petersen P, Fraser C. Strategies to set global analytical quality specificationsin laboratory medicine. Scan J Clin Lab Invest 1999; 59(7): 475–586.
Terrés A. Importancia de la variabilidad biológica y de la relevancia médica. Rev Mex Patol Clin 2003; 50(5): 118-128.
Laurence P, Skendzel M, Barnett R. Medically useful criteria for analytic performance of laboratory test. Am J Clin Pathol 1985; 83(2): 200–205.
Barnett R. Medical significance of laboratory results. Am J Clin Pathol 1968; 50(6): 671- 676.
Simple QC. Tabla de Error máximo permitido CLIA. 1988. (accesado el 8 de mayo de 2014), disponible en (simpleqc. blogspot.mx/p/blogpage_5508.html).
Ricós C, Álvarez V, et al. Desirable specifications for total error, imprecision and Bias, derived from intra e interindividual biologic variation. Accesada el 14 de junio de 2014 disponible en (www.westgard.com).
Cumplido C, Hernández J, Andujo S, Guzmán H. Implementación de Sigma métrico como parte esencial del Control de Calidad Interno en hematología. Rev Latinoam Patol Clin Med Lab 2015; 62(3): 140–145.