Rapid-lecture biological indicators’ monitoring of CEYE (SAPC) autoclaves of the School of Dentistry of the Technological University of Mexico (UNITEC)
Hernández LS, Alavez RS, García HJ, Flores LMG
Language: Spanish
References: 18
Page: 93-97
PDF size: 310.21 Kb.
ABSTRACT
Objective: To monitor autoclave effectiveness and conduct routine validation of sterilization cycles.
Material and methods: An observational, cross-sectioned, qualitative and prospective study of three Sterilization and Packing Center autoclaves of the Technological University of Mexico of the School of Dentistry was conducted. The study included 96 surgical and non-surgical packs, and spanned the period from May 2012 to April 2015. Monthly assessments were conducted, according to Official Mexican Norm stipulation number NOM-013-SSA2-1994, modification 2006, using rapid lecture biological indicators (1292 3M Attest
™). Sterilization efficiency was determined as biological annihilation elicited in these indicators. Effectiveness assessment was conducted after three hours with an automatic optic reader (3M Attest 290
™). A positive witness was included in all readings.
Results: Negative results were obtained in biological indicators readings used in autoclaves. A 100% effectiveness was achieved in sterilization processes of surgical and non surgical packs. All witnesses were positive.
Conclusions: The effectiveness of Sterilization and Packing Center autoclaves of the Technological University of Mexico of the School of Dentistry was corroborated. The Norm established by the Mexican Health Ministry was complied with.
REFERENCES
Troconis-Ganimez JE. Control del ambiente de los consultorios odontológicos: uso de gorro máscara de larga cobertura, bata quirúrgica, dique de goma y guantes. Acta Odontol Venez. 2003; 41 (1): 64-71.
Riera LM, Ambrosio AM. Evaluación de la eficacia de los procesos de esterilización de consultorios odontológicos del Distrito VI de la Provincia de Buenos Aires, Argentina 2006-2007, mediante la utilización de indicadores biológicos. Acta Odontol Venez. 2009; 47 (2): 2-11.
NOM-013-SSA2-2006, para la Prevención y Control de Enfermedades Bucales.
Feria G, Cruz Y. Indicadores biológicos en los procesos de esterilización. Rev Enferm IMSS. 1999; 7 (1): 15-19.
Villalobos SG. Comparación de resultados del proceso de esterilización con dos indicadores biológicos. Rev Enferm IMSS. 2001; 9 (3): 143-146.
Aguirre-Mejía A, Sánchez-Pérez TL. Verificación biológica de los ciclos de esterilización. Rev ADM. 1999; 56 (6): 234-237.
Parra LM, Acosta E. Análisis de algunos indicadores biológicos disponibles comercialmente en México. Rev ADM. 1999; 56 (4): 151-154.
Pardi G, Guilarte C. Algunas consideraciones sobre el control de las infecciones en el consultorio odontológico. Acta Odontol Venez. 2004; 42 (3): 232-233.
Vesley D, Langholz AC, Rohlfing SR, Foltz WE. Fluorimetric detection of a Bacillus stearothermophilus spore-bound enzyme, α-d-glucosidase, for rapid indication of flash sterilization failure. Appl Environ Microbiol. 1992; 58 (2): 717-719.
Albert H, Davies DJ, Woodson LP, Soper CJ. Biological indicators for steam sterilization: characterization of a rapid biological indicator utilizing Bacillus stearothermophilus spore-associated alpha-glucosidase enzyme. J Appl Microbiol. 1998; 85 (5): 865-874.
Perkins RE, Bodman HA, Kundsin RB, Walter CW. Monitoring steam sterilization of surgical instruments: a dilemma. Appl Environ Microbiol. 1981; 42 (2): 383-397.
Gordillo-Vidal ML, Patiño-Suárez MM, Gildo-Medina R. Utilidad en el uso de indicadores biológicos en el proceso de esterilización por calor húmedo. Bioquimia [Internet]. 2007 [consulta 25 de septiembre de 2012]; 32: 118. Disponible en: ISSN 0185-5751
Andrés MT, Tejerina JM, Fierro JF. Reliability of biologic indicators in a mail-return sterilization-monitoring service: a review of 3 years. Quintessence Int. 1995; 26 (12): 865-870.
Acosta-Gío AE, Mata-Portuguez VH, Herrero-Farías A, Sánchez-Pérez L. Biologic monitoring of dental office sterilizers in Mexico. Am J Infect Control. 2002; 30 (3): 153-157.
Reyes CA. Muestreo biológico de autoclaves dentales. Rev Med UV. 2008; 8 (S3): 5-6.
Patiño-Marín N, Loyola-Rodríguez JP. Uso y verificación con indicadores biológicos en esterilizadores de cirujanos dentistas de San Luis Potosí. Salud pública Méx. 2001; 43 (5): 455-458.
Aguilar-Vargas M, Martínez-Jiménez G, Mejía-Hernández JO, Núñez-Morales ZA, Sánchez-León YC, Santos-Huaracha IV. Verificación con indicadores biológicos de equipos esterilizadores (autoclave y calor seco) en las clínicas odontológicas pertenecientes a la FES Iztacala (documento www). Disponible en: http://odontologia.iztacala.unam.mx/memorias15col/contenido/cartel/Verificacionconindicadores05.htm
Osap.org [home page on the internet]. Dentistry’s newsletter for infection control and safety. Infection Control in Practice. 2003 [consulta 12 de octubre de 2012]; 2 (7). Disponible en: http://www.spsmedical.com/documents%26pdfs/steril.monitoring.pdf