2015, Number 3
Ventricular oversensing of atrial electrical activity that inhibits VVI pacemaker and causes syncope
Chávez GE, Alonso HA
Language: Spanish
References: 6
Page: 235-239
PDF size: 386.41 Kb.
ABSTRACT
Far-field oversensing of atrial electrical activity caused by a VVI pacemaker is a rare phenomenon; however, it may have serious clinical consequences. It has several causes and its timely identification may avoid a possible ventricular asystole. This article reports the case of a 72-year-old male who had a Biotronik Axios SR pacemaker implanted, in VVIR mode, six years ago, due to blocked atrial fibrillation. He suffered syncope due to pacemaker inhibition caused by ventricular oversensing of atrial electrical activity.REFERENCES