ABSTRACT
Background: the use of the Program of External Quality Assessment (PEQA) for clinical laboratories is nowadays an indispensable requirement in its accreditation process. The acceptable participation and performance in these programs act as an improvement in the practice of clinical laboratories and instruct about the benefits and limitations of the analytic procedures.
Objective: to demonstrate the usefulness of the assessment of accuracy to detect those laboratories that deviate average consensus, increases straying and cause false results in the evaluation.
Method: a retrospective, longitudinal, descriptive, observational study was conducted to assess accuracy from 1999 to 2013. Samples from the 64 clinical laboratories in the province were processed; in parallel, values included in the level ± 3 standard deviations and rejected (which are out of the level) were considered “accepted” in the reference laboratory of the Coordinating Center of the PEQA.
Results: accuracy improved in time in the laboratories in both levels of attention with the exception of the urate components and leukocyte overall count, that decreased due to difficulties with equipments, reagents or procedures.
Conclusions: during the 14 years of the application of the program in the province, an average of a 74 % of accepted results has been obtained demonstrating that the system proposed is efficient and beneficial.
REFERENCES
García Sánchez M. El laboratorio, piedra angular en el diagnóstico médico. Editorial Asociación Mexicana de Bioquímica Clínica. MÉXICO. Ene-Mar 2006; 32(001):3-5.
Alva Estrada SI, Camacho Aranzúa GY, Escamilla Hernández A. XXX Programa de aseguramiento de la calidad (PACAL). Lab Acta. 2006;18(4):111-5.
Programa de evaluación externa de calidad (PEEC) en el área de Hematología y Hemostasia en diez laboratorios clínicos. Rev Med [Internet]. 2012 [citado 12 Abr 2014];11(2):[aprox. 12 p.]. Disponible en: http://www.uv.mx/rm/num_anteriores/revmedica_vol11_num2/articulos/programa.pdf.
UNE-EN 14136:2004. Utilización de programas de evaluación externa de la calidad en la evaluación del desempeño de los procedimientos de diagnóstico in vitro. Vig. Dic 2004.
Cowley H. Convenio MINSAP - CENIC. Laboratorio de referencia central y control de calidad. Estadígrafos de control de calidad del laboratorio clínico. Actual Lab Clín. 1982;2(3):4-13.
Cuba. Ministerio de Salud Pública. Dirección Nacional de Asistencia Médica. Control de calidad en el laboratorio clínico. Indicación metodológica. Ciudad de La Habana: MINSAP; 1983.
Morejón Campa M, Núñez Santiesteban A, Álvarez Álvarez R. Programa provincial para el control externo de la ca editores. Primera Conferencia Científico –Técnica sobre Aseguramiento de la Calidad; 1989 ene 24 – 26; Ciudad de La Habana. Ciudad de La Habana: Academia de Ciencias de Cuba; 1989. p. 18.
Cuba. Ministerio de Salud Pública. Resolución ministerial no. 215/1996. Anexo: Programa de evaluación externa de la calidad laboratorios clínicos del nivel primario de atención. Ciudad de La Habana: MINSAP; 1996.
Morejón Campa M. El laboratorio clínico y los conceptos asociados a la calidad en Cuba. Diag Vitro [Internet]. 2004 [citado 12 Abr 2014];2:[aprox. 6 p.]. Disponible en: http://www.ifcc.org/ria/div/vol2/morejon.pdf
Morejón Campa M, Ramos Ramos JR, Núñez Santiesteban A. Programa provincial para el control externo de la calidad en los laboratorios clínicos de nivel primario de atención. La Habana: MINSAP; 1998. p. 8-19.
Bethell C. Data and information quality strategic plan. Qual Assur. 2001-2002;9(2):63-97.
Carbajales León AI, Rodríguez Socarras I, López Lastre G. Programa de evaluación externa de la calidad en los laboratorios clínicos. Sus antecedentes y etapa actual en el nivel primario de atención en Camagüey. Hum Méd [Internet]. 2002 Dic [citado 9 Dic 2008];2(4):[aprox. 8 p]. Disponible en:http://www.cmw.sld.cu/amc/v8n3/891.htm.
Rochester MN, Motschaman TL. Corrective and preventive action. Can J Med Technol. 1992;54(2):82-8.
Jenny RW, Jackson Tarentino KY. Causes of unsatisfactory performance in proficiency testing. Clin Chem. 2000;46(1):89-99.
Sierra Amor RI, Hernández Ortiz P, Suárez de Mandujano M, González Solis R. External Quality Scheme Program(EQSP) of the Mexican Association of Clinical Biochemistry. EQA News. 2003;14(3):51-60.
Lawson NS, Glimore BF, Tholen DW. Multiprogram characterization of laboratory bias, precision and total error: proposal for improved assessment with shared external and expanded internal (regional) quality control pools. Arch Pathol Lab Med. 1988;112(4):454-61.
Lawson NS, Glimore BF, Tholen DW. Multiprogram characterization of laboratory bias, precision, and total error. Proposal for improved assessment with shared external and expandednternal (regional) quality control pools. Arch Pathol Lab Med. 1989;112(4):454-81.
Rodríguez Socarras I, Carbajales León AI, Acosta Valdez MA, Hernández Rodríguez J. Programa de evaluación de la veracidad de los análisis de laboratorio en la provincia. Camagüey. Arch Méd Camagüey [Internet]. 2004 [citado 15 Sep 2005];8(3):[aprox. 12 p]. Disponible en: http://www.cmw.sld.cu/amc/v8n3/891.htm
Rodríguez Socarras I, Torres Romo U, Carbajales León AI, Fernández Franch N. Nuestra experiencia en la evaluación externa de la calidad en laboratorios clínicos. Arch Méd Camagüey [Internet].2007 [citado 15 Sep 2005];11(1):[aprox. 2 p]. Disponible en: http://www.amc.sld.cu/amc/2007/v11n1-2007/2170.htm
NC ISO 15189. Laboratorios clínicos Requisitos particulares para la calidad y la competencia. Ciudad de la Habana: Oficina Nacional de Normalización; 2008.
Programa de evaluación externa de calidad (PEEC) en el área de Hematología y Hemostasia en diez laboratorios clínicos. Rev Med [Internet].2013[citado 15 Sep 2005];11(2):[aprox. 2 p].Disponible en: http://www.uv.mx/rm/num_anteriores/revmedica_vol11_num2/articulos/programa.pdf
Acosta Valdez MA, Rodríguez Socarras I. Eficiencia del programa SECAL utilizado en la atención secundaria con la evaluación de la veracidad. Camaguey 1999-2001. Arch Méd Camagüey [Internet]. 2003 [citado 15 sep 2005];7(Supl 2): [aprox. 12 p]. Disponible en: http://www.amc.sld.cu/amc/2003/v7supl2/782.htm
OGA-GEC -006. Criterios para la acreditación de laboratorios de ensayo y calibración. Guatemala: Oficina de Acreditación Guatemala; 2003