Dislocation of the constrained implant in total hip arthroplasty. A case report
Díaz-Borjón E, Lora-Martínez M, Montejo-Vargas J, Astudillo-Mata P
Language: Spanish
References: 17
Page: 25-28
PDF size: 515.20 Kb.
ABSTRACT
Constrained acetabular components are seldomly used for the treatment of recurrent hip
prosthesis dislocation. We present the case of a 48 female patient with recurrent hip
arthroplasty treated with a constrained total hip replacement. She sustained 3
additional dislocation episodes of the constrained implant before being revised
to a standard non-constrained cup. All three dislocations occurred with the femoral
head being disengaged from the constrained polyethylene cup. Revision of the
literature reveals that this dislocation fashion has been reported in only
three previous cases, with a disloged cup or polyethylene insert being the
most common presentation.
REFERENCES
Anderson MJ, Murray WR, Skinner HB: Constrained acetabular components. J Arthroplasty 1994; 9(1): 17-23.
Bryan WJ, Reeve RE: Dislocation and failure of an articulated total hip replacement. A case report. Orthopedics 1986; 9(8): 1113-5.
Cameron HU. Use of a distal trochanteric osteotomy in revision hip surgery. Contemp Orthop 1991; 23(5): 235-238.
Dorr LD, Wolf AW, Chandler R, Conaty JP: Classification and treatment of dislocations of total hip arthroplasty. Clin Orthop 1983; (173): 151-8.
Fackler CD, Poss R: Dislocation in total hip arthroplasties. Clin Orthop 1980; (151): 169-78.
Goetz DD, Capello WN, Callaghan JJ, Brown TD, Johnston RC: Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A retrospective analysis of fifty-six cases. J Bone Joint Surg 1998; 80(4): 502-509.
Kaper BP, Bernini PM: Failure of a constrained acetabular prosthesis of a total hip arthroplasty. A report of four cases. J Bone Joint Surg 1998; 80(4): 561-5.
Kavanagh BF, Fitzgerald RH Jr: Multiple revisions for failed total hip arthroplasty not associated with infection. J Bone Joint Surg 1987; 69(8): 1144-9.
Koffman M: Proximal femoral resection or total hip replacement in severely disabled cerebral-spastic patients. Orthop Clin North Am 1981; 12(1): 91-100.
Lombardi AV Jr, Mallory TH, Kraus TJ, Vaughn BK: Preliminary report on the S-ROM constraining acetabular insert: A retrospective clinical experience. Orthopedics 1991; 14(3): 297-303.
Morrey BF: Joint Replacement Arthroplasty, New York, Churchill Livingstone, 1991, p. 861.
Olerud S, Karlstrom G: Recurrent dislocation alter total hip replacement. Treatment by fixing an additional sector to the acetabular component. J Bone Joint Surg 1985; 67(3): 402-5.
Paterno SA, Lachiewickz PF, Kelley SS: The influence of patient-related factors and the position of the acetabular component on the rate of dislocation after total hip replacement. J Bone Joint Surg 1997; 79(8): 1202-10.
Radulovic B, Kenig I, Radovanovic M: Indications for Sivash type total hip prosthesis, in Charnley J (ed): Low Friction Arthroplasty of the Hip: Theory and Practice. Berlin, Germany: Springer-Verlag, 1979, pp. 74-81.
Russin LA, Russin MA. Abstract: A preliminary study of total hip joint replacement by jeans of the Russin-modified Sivash prostheses: 100 cases. 41st Annual Meeting Proceedings, Dallas, Texas. Chicago, IL: American Academy of Orthopaedic Surgeons, 1974, p. 119.
Russin LA, Sonni A: Treatment of intertrochanteric and subtrochanteric fractures with Ender’s intramedullary rods. Clin Orthop 1980; (148): 203-212.
Sivash KM: The development of a total metal prosthesis for the hip joint from a partial joint replacement. Reconstr Surg Traumatol 1969; 11: 53-62.