Table 3: The different measurements that microvascular clamps can have, the pressure they can exert and their relationship with the diameter of the blood vessels are shown. |
|||||||
Clamp Serie |
a (mm) |
b (mm) |
e (mm) |
d (mm) |
Pressure in grams |
Pressure in grams |
Blood vessel diameter (mm) |
RD |
6.0 |
36.0 |
16.0 |
3.0 |
120-140 |
– |
2.0-5.0 |
HD |
4.0 |
24.0 |
10.0 |
2.0 |
60-75 |
30-40 |
1.5-3.5 |
B-3 |
3.5 |
16.6 |
7.5 |
1.7 |
30-40 |
– |
1.0-2.2 |
B-2 |
3.0 |
11.3 |
5.5 |
1.5 |
17-23 |
– |
0.6-1.5 |
B-1 |
2.2 |
8.0 |
3.8 |
1.1 |
10-14 |
– |
0.4-1.0 |