Cuba y Salud 2018; 13 (2)
Assessment of electro chemistry method for International Normalization Reason determination
Cala FJ, Gondres LKM, Robert TV, Delisle BX, Naranjo LY, Rodríguez BS
Language: Spanish
References: 15
Page: 6-10
PDF size: 1081.63 Kb.
ABSTRACT
Objectives: to evaluate the usefulness of Coaguchek XS Pro to determine the Protombine / INR time and
compare the values of the International Normalized Ratio obtained by electrochemical method, in correspondence
with the reference method.
Methods: a descriptive study of 420 patients with adjusted doses of oral anticoagulants was carried out in the
haemostasis laboratory of the Clinical Laboratory Department of the “Saturnino Lora Torres” Provincial Hospital
in Santiago de Cuba, from May to June. 2016. Student’s t-test and correlation measures were applied.
Results: from the total of samples 388 cases 93.2% coincided for a high level of significance and a Pearson
correlation coefficient of 0.93 was obtained between the electrochemical and electromagnetic methods.
Conclusions: it is concluded that the determination of the International Normalized Ratio using Coaguchek XS
Pro is satisfactory compared to the standard method in the therapeutic range, although the higher the value of
the International Normalized Ratio, the greater the differences between them.
REFERENCES
Rodríguez Borges S, Durañones Góngora S, Castillo Martínez JM, Romero García LI, Gondres Legró KM. Variaciones de la Razón Normalizada Internacional en pacientes con prótesis valvulares complicados. MEDISAN 2017; 21(4):460. Disponible en: http:// scielo.sld.cu/scielo.php
Coagulómetros portátiles. Revisión de la evidencia científica y evaluación económica de su uso en el autocontrol del tratamiento anticoagulante oral. IN06/2003. Disponible en : www.etsad.fr/etsad/afficher_lien.php
Duboscq C, Ceresetto JM, Stemmelin G, Shanley C, Rabinovich O, Palmer S et al. Evaluación del desempeño de un dispositivo portátil para la determinación de la razón internacional normatizada en una clínica de anticoagulación. HEMATOLOGÍA. 2014;18(3):204-210. Disponible en: http://www.sah.org.ar/revista/numeros/vol.18n.3.204.210.pdf
CoaguChek XS. El coagulómetro portátil de nueva generación para el autocontrol de INR. [Artículo en línea]. 2016. Disponible en: http://www.coaguchek.com/resource.php
Candela Thoa AM. Validación de aparatos y métodos de medida: concordancia sí, correlación no. MedClin (Barc). Disponible en : www.elsevier.es/es-revista-semergen-medicina-familia-40-pdf-S1138359301740130-S300
Gregory C, Sephel, Laposata M. A Case Study Demonstrating the Need for Ongoing Correlation of POC With the Central Laboratory. Am J Clin Pathol. 2013; 140:475-486. Disponible en : https://www.ncbi.nlm.nih.gov/pubmed/24045543
Beynon C, Jakobs M, Rizos T, Unterberg AW, Sakowitz OW. Rapid bedside coagulometry prior to urgent neurosurgical procedures in anticoagulated patients. Br J Neurosurg. 2014;28(1):29-33. Disponible en: https://www.ncbi.nlm.nih.gov/pubmed/24313307
Consenso: VII Conferencia de Terapia Antitrombótica del American Collage of Chest Physicians (ACCP). 2004.
Kalcık M, Yesin M, OzanGu¨ rsoy M, Gu¨ndu¨ z S, Karakoyun S, Astarcıoglu M et al. Comparison of the INR Values Measured by CoaguChek XS Coagulometer and Conventional Laboratory Methods in Patients on VKA Therapy. Clinical and Applied Thrombosis/ Hemostasis. 2016 Disponible en : https://www.researchgate.net/
Dawn E Havrda, Toni L Hawk, Carrie M Marvin. Accuracy and Precision of the CoaguChek S versus Laboratory INRs in a Clinic. Ann Pharmacother. 2002;36(5): 769-775. Disponible: https://www.ncbi.nlm.nih.gov/pubmed/11978150
Donaldson M, Sullivan J, Norbeck A. Comparison of international normalized ratios provided by two point-of-care devices and laboratory-based venipuncture in a pharmacist-managed anticoagulation clinic. Am J Health-Syst Pharm. 2010;67(19): 1616- 1622. Disponible en : www.medscape.com/viewarticle/
Duboscq C, Ujhelly C, Mainetti G. Estudio Comparativo multicéntrico del RIN en Coagucheck XS vs equipos automatizados. XI Congreso del Grupo Argentino de Hemostasia. [Poster Nro 37]. Argentina.2014
Hur M, Kim H, Park CM, LaGioia A, Choi SG, Choi JH et al. Comparison of international normalized ratio measurement between CoaguChek XS Plus and STA-R coagulation analyzers. Biomed Res Int. 2013. Disponible en: http://biblioteca.universia.net/html_ bura/ficha/params/title/comparison-of-international-normalized-ratio-measurement-between-coaguchek-xs-plus/id/64023298. html
Myung-Hyun N, Kyoung Ho R, Hui-Nam P, Chang Kyu L, Young-Hoon K, Kap No L, et al. Evaluation of the Roche CoaguChek XS Handheld Coagulation Analyzer in a Cardiac Outpatient Clinic. Ann Clin Lab Sci Winter. 2008;38(1):37-40.Disponible en: https://www.ncbi.nlm.nih.gov/pubmed/18316780
Caballero-Villarraso J, Villegas-Portero R, Rodríguez-Cantalejo F. Dispositivos de coagulometría portátil en el seguimiento y control ambulatorio de la terapia anticoagulante oral: revisión sistemática. Aten Primaria. 2011; 43(3):148-156. Disponible en : www.elsevier.es