Three successive fractures of different hip femoral stems on the same patient
Gómez-Robledo J
Language: Spanish
References: 11
Page: 138-143
PDF size: 298.01 Kb.
ABSTRACT
The clinical case of an 80-year-old female who suffered three fractures in uncemented stems implanted in her right hip in a period of three years is presented. A fracture occurred in the prosthetic neck and the other two at the juncture of the conical and cylindrical stem portion, coinciding with the metaphyseal-diaphyseal junction of the femur. The main causes of the failure were an increasing concentration of forces at the level of the implant as a consequence of the increased length of the neck and lateralization of the femur (offset). Other causes that have contributed to this prosthetic failure are analyzed.
REFERENCES
Bonete-LLuck D, Pulido-Higueras MC, Maroñas-Abuelo C, Guinot-Tormo JM, Pardo-Montaner J: Rotura de vástago femoral de prótesis total de cadera enteramente recubierta de hidrixiapatita. Presentación de tres casos. Rev Esp Cir Osteoart. 2007; 42(229): 14-8.
Sharma DK, Brooks S, Creasey S, Lewis B: Fracture of fully hydroxyapatite-coated titanium femoral stem of a total hip replacement--a report of 3 cases. Acta Orthop Scand. 2004; 75(6): 768-71.
Harvie P, Haroon M, Henderson N, El-Guindi M: Fracture of the hydroxyapatite-ceramic-coated JRI-Furlong femoral component. J Bone Joint Surg (Br). 2007; 89-B: 742-5.
Martens M, Aernoudt E, de Meester P, Ducheyne P, Mulier JC, de Langh R, et al: Factors in the mechanical failure of the femoral component in total hip prosthesis. Report of six fatigue fractures of the femoral stem and results of experimental loading tests. Acta Orthop Scand. 1974; 45(5): 693-710.
Garg B, Mittal R, Rastogi S: Femoral prosthesis neck fracture following total hip arthroplasty: a case report. Acta Orthop Belg. 2011; 77(3): 406-9.
Morgan-Hough CV, Tavakkolizadeh A, Purkayastha S: Fatigue failure of the femoral component of a cementless total hip arthroplasty. J Arthroplasty. 2004; 19(5): 658-60.
Sangüesa-Nebot MJ, Soriano FC, Gabarda RF, Mordt CV: Revision hip arthroplasty with a short femoral component in fractured hydroxyapatite fully coated femoral stem. J Arthroplasty. 2010; 25(7): 1168.e13-6.
Yoshimoto K, Nakashima Y, Nakamura A, Mawatari T, Todo M, Hara D, et al: Neck fracture of femoral stems with a sharp slot at the neck: biomechanical analysis. J Orthop Sci. 2015; 20(5): 881-7.
Furlong® H-A.C. Revision Hip Stems A.C. & Securus Revision Hip System 155-030 Issue 6 - June 2012.
Trikha SP, Singh S, Raynham OW, Lewis JC, Mitchell PA, Edge AJ: Hydroxyapatite-ceramic-coated femoral stems in revision hip surgery. J Bone Joint Surg Br. 2005; 87(8): 1055-60.
Vatani N, Comando D, Acuña J, Prieto D, Caviglia H: Faulty design increases the risk of neck fracture in a hip prosthesis. Acta Orthop Scand. 2002; 73(5): 513-7.