Comparación de los valores del espesor corneal central según los equipos Lenstar, Galilei y Pentacam
Miranda HI, Fernández GK, Ruiz RY, Hernández SJR, Rio TM
Language: Spanish
References: 15
Page:
PDF size: 127.43 Kb.
ABSTRACT
Objective: To compare the central corneal thickness measurements taken by pachimetry with the partial coherence interferometry, Lenstar and with Scheimpflug systems (Pentacam; Oculus) and Galilei (Ziemer, Switzerland).
Methods: Comparative and prospective study of 120 eyes from 60 patients. Several recommended measurements were taken with the optical biometers LenstarLS 900 (Haag Streit AG) and with Pentacam y Galilei topographers. The results were evaluated using the linear regresión analysis and Pearson´s correlation.
Results: There is high correlation among the central corneal thickness measurements taken by the Lenstar equipment and topographers Pentacam and Galilei.
Conclusion: The new biometer LENSTAR provided results that correlated very well with those of the Pentacam and Galilei systems. The Lenstar is a precise device that will be helpful for any cataract or refractive surgery.
REFERENCES
Sanctis U, Missolungi A, Mutani B, Richiardi L, Grignolo FM. Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol. 2007 Nov;144(5):712-8.
Colín Malagón A. Exploración preoperatoria de la catarata: Plan quirúrgico. ¿Cuál es el objetivo de la cirugía? En: Alió JL. Buscando la excelencia en la Cirugía de Catarata. Barcelona: Editorial Glosa; 2007. p. 28-39.
Concepción AL. Tonometría: Actualidades y novedades. [Internet]. Panama: Jaypee Highlights Medical Publisher; 2007 [citado 12 enero de 2009]. Disponible en: http://www.elibraryweb.net/content/view/687/9/
González MJ. Cirugía refractiva en pacientes con glaucoma. Gabinete tecnológico del glaucoma. [actualizado 2005; citado 12 enero de 2009]. Disponible en: http://www.mjgonzakezorg/cms/index.php?id=3.29.0.0.1.0
Rabsilber TM, Khoramnia R, Auffart GU. Anterior chamber measurements using Pentacam rotating Scheimpflug camera. J Cataract Refract Surg. 2006;32(3):456-9.
Lackner B, Schmidinger G, Pieh S, Funovics MA, Skorpik C. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci. 2005 Oct;82(10):892-9.
Buckhurst PJ, Wolffsohn JS, Shah S, Naroo SA, Davies LN, Berow EJ. A new optical low coherence reflectometry device for ocular biometry in cataract patients. Br J Ophthalmology. 2009;93(7):949-53.
Drexler W, Findl O, Menapace R, Rainer G, Vass C, Hitzenberger CK, et al. Partial coherence interferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol. 2003;126(4):524-34.
Santodomingo-Rubido J, Mallen EA, Gilmartin B, Wolffsohn JS. A new non-contact optical device for ocular biometry. Br J Ophthalmol. 2002;86(4):458_62.
Böhnke M, Widmer S, Wälti R. Real-time pachymetry during photorefractive keratectomy using optical low-coherence reflectometry. J Biomed Opt. 2001;6(4):412_7.
Schmid GF. Axial and peripheral eye length measured with optical low coherence reflectometry. J Biomed Opt. 2003;8(4):655-8.
Bohnke M, Master BR, Walti R, Ballif JJ, Chavanne P, Gianotti R, et al. Precision and reproducibility of meansurements of human cornela thickness with rapid optical low-coherence interferometry (OLCR). J Biomed Opt. 1999;4(1):152-6.
Ucakhan ÖÖ, Özkan M, Kanpolat A. Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry. J Cataract Refract Surg. 2006;32(6):970-7.
Barkana Y, Gerber Y, Elbaz U, Shwartz S, Ken-Dror G, Ovni I, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg. 2005;31(9):1729-35.
Paredes García B, Carrasco Font C, Arias Fuentes A, García Sáenz MC. Relación entre el espesor corneal central y la medida de la presión intraocular con tonometría de aplanación. Microcirugía Ocular. 2006 [citado 12 enero de 2009];(3). Disponible en: http://www.oftalmo.com/secoir2006/rev06-3/06c-07.htm