Evaluation of the application of the UMELISA HIV 1+2 RECOMBINANT in samples of dry blood on filter paper S & S 903
Morales-Rodríguez E, Alonso-Ramos V, Pozo-Peña L, González-Suárez A, Figueiras-Lache Y, González-Quintero A, López-Brauet L, Alvarez-García M, Turró-Grau G, Rizo-Polo D, Gómez-Cordero I
Language: Spanish
References: 15
Page: 149-155
PDF size: 45.10 Kb.
ABSTRACT
318 samples from supposedly healthy blood donors at the blood bank were studied. As a result, 315 negative samples, two borderline samples, and a positive one were obtained. A DAVIH-BLOT test confirmed the positive, negative and borderline sample obtained by UMELISA. A 99.68% specificity was obtained with this test. With the same aim in mind, 233 positive samples in SH, FP S & S 903 to 45% and 55% hematocrit were studied. They maintained the same levels of positivity with both methods, and 100% sensitibity to both values of hematocrit.
REFERENCES
Desjarlis DC. Antibodies to a retrovirus etiologically associated with acquired inmunodeficiency syndrome (AIDS) in populations with increased incidence of the syndrome. MMWR 1984; 33:377-9.
Ziegler JB. Postnatal transmission of AIDS associated retrovirus from mother to infant. Lancet 1985; 1:896-7.
Gallo RC. Frecuent detection and isolation of cythopathic retroviruses (HTLV-I/II) from patient with AIDS and at risk for AIDS. Science 1984; 224:500-3.
Feinberb MA and Volberding PA. Testing for Human Immunodeficiency Virus. In: Cohen PT, Sande MA, and Volberding PA. The Aids Knowledge Base Boston: Little, Brown and Company, 1994: Section 2.
Roy, R Middendorf. Infrared fluorescent detection of D1S80 alleles from blood and body fluid collected on IsoCode devices. Biotechniques 1997; 23, 942-5.
Chanbancherd P, Brown AE, Trichavaroj R, Tienamporn P, Puthakird P, et al. Application of dried blood spot specimens for serologic subtyping of human immunodeficiency virus type 1 in Thailand. J Clin Microb 1999; 30: 804-6.
Behets F, Kashamuka M, Pappaioanou M, Green TA, Ryder R, et al. Stabilitiy of human immunodeficiency virus type 1 antibodies in whole blood dried on filter paper and stored under various tropical conditions in Kinshasa, Zaire. J Clin Microb 1992; 30:1179-82.
Van den akker B. Recovery of HIV antibodies in eluates from plasma and erythrocytes dried on filter paper and stored under various conditions. AIDS 1990; 4:87-92.
Evengaard B.Effect of heat on extracted HIV infectivity and antibody activity using the filter paper technique of blood sampling. AIDS 1989; 3:391-5.
Novoa LI. Cloning and expression of the major core HIV-1 protein in bacteria. 1er Congreso Iberoamericano de Biotecnología. Libro de Resúmenes (s05-051). La Habana, Cuba, 1989.
Novoa LI. Development of a system for HIV seroconversion using two recombinant antigens. International Biotechnology Congress. Paris, 1988.
Silva C. UMELISA con antígeno recombinante para el pesquisaje de anticuerpos Anti-VIH. III Reunión Iberolatinoaméricana de Hematología. Libro de Resúmenes, La Habana, Cuba. 1989. p. 203.
Zady M. Correlation and simple least squares regression. Clinical Laboratory Science Program, University of Louisville. Louisville, Kentucky; 2000.
World Health Organization. Proposed WHO criteria for interpreting results from western blot assays for HIV-1, HIV-2 and HTLV-I/II. Wkly Epidemiol Rec, 1990; 65:283-8.
Weiss SH. Laboratory detection of human retroviral infection. In: Wormser GP. AIDS and Other Manifestations of HIV Infectation. New York: Lippincott-Raven;1998. p. 175-200.