2002, Number 3
<< Back Next >>
salud publica mex 2002; 44 (3)
Hemoglobin measured by Hemocue and a reference method in venous and capillary blood: A validation study
Neufeld L, García-Guerra A, Sánchez-Francia D, Newton-Sánchez O, Ramírez-Villalobos MD, Rivera-Dommarco J
Language: English
References: 27
Page: 219-227
PDF size: 61.94 Kb.
ABSTRACT
Objective. To assess the comparability of hemoglobin concentration (Hb) in venous and capillary blood measured by Hemocue and an automated spectrophotometer (Celldyn) and to document the influence of type of blood (capillary or venous) and analysis method on anemia prevalence estimates.
Material and Methods. Between February and May 2000, capillary and venous samples were collected from 72 adults and children at Hospital del Niño Morelense (Morelos State Children’s Hospital) in Cuernavaca, Morelos, Mexico, and assessed for Hb using the Hemocue and Celldyn methods. Estimated Hb levels were compared using the concordance correlation coefficient and Student’s t test for paired data. The sensitivity and specificity for anemia diagnosis were estimated and compared between type of blood and method of assessment.
Results. Capillary blood had higher Hb (+0.5g/dl) than venous blood in adults and children, as did samples assessed by Celldyn compared to Hemocue (+0.3g/dl). Specificity to detect anemia was adequate (›0.90) but sensitivity was low for capillary blood assessed by Hemocue (‹0.80).
Conclusions. The difference in Hb between venous and capillary blood is likely related to biological variability. Hemoglobin concentration in capillary blood assessed by Hemocue provides an adequate estimation of population anemia prevalence but may result in excess false negative diagnoses among individuals. The results of this study stress the importance of sample collection technique, particularly for children. Method of analysis and sampling site need to be taken into consideration in field studies.
REFERENCES
DeMaeyer E, Adiels-Tegman M. The prevalence of anemia in the world. World Health Stat Q 1985;38:302-316.
Administrative Committee on Coordination, Sub-Committee on Nutrition. (ACC/SCN) Third report on the world nutrition situation. Ginebra: United Nations, 1997.
Morris S, Ruel M, Cohen R, Dewey K, Brière BDL, Hassan M. Precision, accuracy, and reliability of hemoglobin assessment with use of capillary blood. Am J Clin Nutr 1999;69:1243-1248.
Daae LN, Halvorsen S, Maticen PM, Mironska K. A comparison between haematological parameters in ‘capillary’ and venous blood from healthy adults. Scand J Clin Lab Invest 1988;48:723-726.
Hutler M, Beneke R, Boning D. Determination of circulating hemoglobin mass and related quantities by using capillary blood. Med Sci Sports Exerc 2000;32:1024-1027.
Moe PJ. Hemoglobin, hematocrit and red blood cell count in “capillary” (skin-prick) blood compared to venous blood in children. Acta Paediatr Scand 1970;59:49-51.
Thomas WJ, Collins TM. Comparison of venipuncture blood counts with microcapillary measurements in screening for anemia in one-year-old infants. J Pediatr 1982;101:32-35.
Bridges N, Parvin RM, van Assendelft OW. Evaluation of a new system for hemoglobin measurement. Am Clin Products Rev 1987;6:22-25.
Cohen A, Seidl-Friedman J. HemoCue system for hemoglobin measurement. Am J Clin Pathol 1988;90: 302-305.
Neville R. Evaluation of portable haemoglobinometer in general practice. Br Med J 1987;294:1263-1265.
Chen P, Short T, Leung D, Oh T. A clinical evaluation of the Hemocue haemoglobinometer using capillary, venous, and arterial samples. Anaesth Intens Care 1992;20:497-503.
Statlande BE, Winkel P, Harris SC, Burdsall MJ, Saunders AM. Evaluation of biological sources of variation of leukocyte counts and other hematologic quantities using very precise automated analyzers. Am J Clin Pathol 1977;69:48-54.
Cohen J. Statistical power analysis for the behavioral sciences. Second edition. Nueva Jersey (NJ): Hillsdale Publishers, 1988.
von Schenck H, Falkensson M, Lundberg B. Evaluation of “HemoCue”, a new device for determining hemoglobin. Clin Chem 1986;32:526-529.
Drabkin DL, Austin JH. Spectrophotometric studies: Spectrophotometric constants for common hemoglobin derivatives in human, dog and rabbit blood. J Biol Chem 1932;98:719-733.
Bigbee J, Winter B. Reliability and validity testing of the whole-blood hemoglobinometer. Public Health Nurs 1990;7:40-44.
Sandberg S, Thue G, Christensen HG, Lund PK, Rynning M. Performance of cell counters in primary health care. Scand J Prim Health Care 1991;9:129-133.
Petani S, Topic E, Turcic G, Daschner M. Clinical evaluation of the Cell- Dyn 1700CS blood counter. Clin Chem 1997;43:1085-1088.
Sachse C, Henkel E. An evaluation of the CELL-DYN 1700 haematology analyzer: Automated cell counting and three-part leucocyte differentiation. Clin Lab Haematol 1996;18:171-180.
Marks GC, Habicht JP, Mueller WH. Reliability, dependability, and precision of anthropometric measurements. The Second National Health and Nutrition Survey 1976-1980. Am J Epidemiol 1989;130:578-587.
Lin L. A concordance correlation coefficient to evaluate reproducibility. Biometrics 1989;45:255-268.
Lin L. Assay validation using the concordance correlation coefficient. Biometrics 1992;48:599-604.
Bland JM, Atlman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;8176:307-310.
Gillespie S, Johnston JL. Expert consultation on anemia determinants and interventions. Ottawa: Micronutrients Initiatives, 1998.
Cohen JH, Haas JD. Hemoglobin correction factors for estimating the prevalence of iron deficiency anemia in pregnant women residing at high altitudes in Bolivia. Rev Panam Salud Publica 1999;6:392-399.
Ruiz-Argüelles GJ, Llorente-Peters A. Predictive equations for normal red cell values at altitudes from sea level to 2,670 meters. Rev Invest Clin Mex 1981;33:191-193.
Freire WB. Strategies of the Pan American Health Organization for the control of iron deficiency in Latin America. Nutr Rev 1997;55:183-188.