2020, Number 4
Accuracy of implant impression using a multifunctional guide in the All-on-Four treatment concept
Language: Spanish
References: 30
Page:
PDF size: 380.27 Kb.
ABSTRACT
Introduction: The passivity of prosthetic components is one of the most important prerequisites in oral implant rehabilitation for maintenance of osseointegration.Objective: Thus, the present study analyzed in vitro the accuracy of different molding techniques in prosthetic rehabilitation on angled implants installed in an anatomical model of a metallic edentulous maxilla.
Methods: Laboratory experiment study. A reference metal model of an edentulous maxilla was used. A metal framework was used for the misadaptation assessment. Three groups (n= 10) were compared and impressions were made with vinyl-polysiloxane and casts were obtained with type IV stone: (1) Impression with a metallic open-tray without splinted transfers; (2) Impression with a metallic open-tray and metal splinted transfers; (3) Impression with a multifunctional guide and metal splinted transfers. Misadaptation was assessed with the aid of a stereomicroscope and measuring software. Data were submitted to ANOVA with Welch correction and the Games-Howell post-hoc test with the significance set at 5 %.
Results: Misadaptation (µm) was 110,23 ± 30,94, 37,53 ± 3,92, and 37,69 ± 2,79 for the the groups. Statistically significant differences between impression with a metallic open-tray with and without splinted transfers were observed (p < 0,001). No significant differences between the other types of impression.
Conclusions: Impression with a metallic open-tray without splinted transfers and with a multifunctional guide splinted with metal transfers were the most precise methods resulting in higher accuracy in transferred implants.
REFERENCES
Gupta S, Narayan AI, Balakrishnan D. In Vitro Comparative Evaluation of Different Types of Impression Trays and Impression Materials on the Accuracy of Open Tray Implant Impressions: A Pilot Study. Int J Dent. 2017 [access: 10/01/2019]; 2017:1-8. Available from: Available from: https://www.hindawi.com/journals/ijd/2017/6306530/
Silva GC, Mendonça JA, Lopes LR, Landre J. Stress patterns on implants in prostheses supported by four or six implants: a three-dimensional finite element analysis. Int J Oral Maxillofac Implants. 2010 [access: 10/01/2019]; 25(2):239-46. Available from: Available from: http://www.ncbi.nlm.nih.gov/pubmed/20369081
Dellavia C, Francetti L, Rosati R, Corbella S, Ferrario VF, Sforza C. Electromyographic assessment of jaw muscles in patients with All-on-Four fixed implant-supported prostheses. J Oral Rehabil. 2012 [access: 10/01/2019]; 39(12):896-904. Available from: Available from: http://doi.wiley.com/10.1111/joor.12002
de Avila ED, Barros LAB, Del’Acqua MA, Castanharo SM, Mollo FA. Comparison of the accuracy for three dental impression techniques and index: An in vitro study. J Prosthodont Res. 2013 [access: 12/01/2019]; 57(4):268-74. Available from: Available from: https://linkinghub.elsevier.com/retrieve/pii/S188319581300090X
Manzella C, Bignardi C, Burello V, Carossa S, Schierano G. Method to improve passive fit of frameworks on implant-supported prostheses: An in vitro study. J Prosthet Dent. 2016 [access: 12/01/2019]; 116(1):52-8. Available from: Available from: https://www.sciencedirect.com/science/article/pii/S0022391316000512
Geramipanah F, Sahebi M, Davari M, Hajimahmoudi M, Rakhshan V. Effects of impression levels and trays on the accuracy of impressions taken from angulated implants. Clin Oral Implants Res. 2015 [access: 17/01/2019]; 26(9):1098-105. Available from: Available from: http://doi.wiley.com/10.1111/clr.12410
Papaspyridakos P, Hirayama H, Chen CJ, Ho CH, Chronopoulos V, Weber HP. Full-arch implant fixed prostheses: a comparative study on the effect of connection type and impression technique on accuracy of fit. Clin Oral Implants Res . 2016 [access: 20/01/2019]; 27(9):1099-105. Available from: Available from: http://doi.wiley.com/10.1111/clr.12695
Mpikos P, Kafantaris N, Tortopidis D, Galanis C, Kaisarlis G, Koidis P. The effect of impression technique and implant angulation on the impression accuracy of external- and internal-connection implants. Int J Oral Maxillofac Implants . 2012 [access: 20/01/2019]; 27(6):1422-8. Available from: Available from: http://www.ncbi.nlm.nih.gov/pubmed/23189292
Del’Acqua MA, Arioli-Filho JN, Compagnoni MA, de Assis Mollo Jr. F . Accuracy of Impression and Pouring Techniques for an Implant-Supported Prosthesis. Int J Oral Maxillofac Implants . 2008 [access: 11/02/2019]; 23(2):226-36. Available from: Available from: http://www.ncbi.nlm.nih.gov/pubmed/18548918
Del’acqua MA, de Avila ÉD, Amaral ÂLC, Pinelli LAP, Assis-Mollo F. Comparison of the accuracy of plastic and metal stock trays for implant impressions. Int J Oral Maxillofac Implants . 2012 [access: 15/02/2019]; 27(3):544-50. Available from: Available from: http://www.ncbi.nlm.nih.gov/pubmed/22616047
Alexander Hazboun GB, Masri R, Romberg E, Kempler J, Driscoll CF. Effect of implant angulation and impression technique on impressions of NobelActive implants. J Prosthet Dent . 2015 [access: 10/03/2019]; 113(5):425-31. Available from: Available from: https://www.sciencedirect.com/science/article/pii/S0022391314005423
Shim JS, Ryu JJ, Shin SW, Lee JY. Effects of Implant Angulation and Impression Coping Type on the Dimensional Accuracy of Impressions. Implant Dent . 2015 [access: 10/03/2019]; 24(6):726-9. Available from: Available from: http://content.wkhealth.com/linkback/openurl?sid=WKPTLP:landingpage&an=00008505-900000000-99556
Rutkunas V, Sveikata K, Savickas R. Effects of implant angulation, material selection, and impression technique on impression accuracy: a preliminary laboratory study. Int J Prosthodont . 2012 [access: 05/04/2019]; 25(5):512-5. Available from: Available from: http://www.ncbi.nlm.nih.gov/pubmed/22930776
Pujari M, Garg P, Prithviraj DR. Evaluation of Accuracy of Casts of Multiple Internal Connection Implant Prosthesis Obtained From Different Impression Materials and Techniques: An In Vitro Study. J Oral Implantol. 2014 [access: 12/04/2019]; 40(2):137-45. Available from: Available from: http://www.ncbi.nlm.nih.gov/pubmed/24456531
Hariharan R, Shankar C, Rajan M, Baig MR, Azhagarasan NS. Evaluation of accuracy of multiple dental implant impressions using various splinting materials. Int J Oral Maxillofac Implants . 2010 [access: 13/04/2019]; 25(1):38-44. Available from: Available from: http://www.ncbi.nlm.nih.gov/pubmed/20209185
Tsagkalidis G, Tortopidis D, Mpikos P, Kaisarlis G, Koidis P. Accuracy of 3 different impression techniques for internal connection angulated implants. J Prosthet Dent . 2015 [access: 13/04/2019]; 114(4):517-23. Available from: Available from: https://www.sciencedirect.com/science/article/pii/S0022391315002826
de Avila ED, de Matos Moraes F, Castanharo SM, Del’Acqua MA, de Assis Mollo F. Effect of Splinting in Accuracy of Two Implant Impression Techniques. J Oral Implantol . 2014 [access: 2019 Apr 20]; 40(6):633-9. Available from: Available from: http://www.joionline.org/doi/abs/10.1563/AAID-JOI-D-12-00198
Matta R-E, Bergauer B, Adler W, Wichmann M, Nickenig H-J. The impact of the fabrication method on the three-dimensional accuracy of an implant surgery template. J Cranio-Maxillofacial Surg. 2017 [access: 23/04/2019]; 45(6):804-8. Available from: Available from: https://www.sciencedirect.com/science/article/abs/pii/S1010518217300756